Description/ Specification of AFS Lamp
AFS lamp KY-AF In order to gain good sensitivity in detection with atomic fluorescence spectrophotometer, the signal-to-noise ratio is dependent on following three aspects: Fluorescent lamp of instrument light source Atomization sampling system Instrument acquisition and processing system Element As ﹒ Bi ﹒ Cd ﹒ Ge ﹒ Hg ﹒ Pb ﹒ Au ﹒ Cr and Cu Element Sb ﹒ Se ﹒ Sn ﹒ Te ﹒ Zn ﹒ Ni ﹒ Fe ﹒ Co and Ag ﹒ Hydride process Others adopt flame method Fluorescent lamp stability: JJG standard:<2% (preheat for 30min and record for 30min) Manufacturing standard: ≤ 1.5% (preheat for 30min, record 30min) Energy of fluorescent lamp: Hg:Standard solution 1ng, reading value of instrument>1500 As:Standard solution 10ng, reading value of instrument > 2000 AFS lamp KY-AF-A Stability of AFS lamp: ≤ 1.5% (preheat for 30min and record for 30min) Emitted energy of AFS lamp: Hg standard solution 1Reading value of 1ng instrument >1500 As standard solution Reading value of 10ng instrument >2000 Concentricity of ASF light: When the light rotates 360 ° , the shift to reading value of instrument is <15%